Self-Contained Biological Indicators for monitoring steam sterilization consist of: ●A plastic pipe and cap. The cap with a hole is placed on top of the pipe to allow the sterilizing agent to penetrate into the indicator. Between pipe and cap there is a filter paper which is used to avoid pollution. ●A crushable media ampule which contains modified culture medium with a pH indicator and fluorogenic substrates. The modified culture medium will transition from the initial Purple color to Yellow in the presence of bacterial growth. ●An inoculated carrier of Geobacillus stearothermophilus ATCC 7953 with a population level of ≥106 is located at the bottom closed end of the pipe. ●A barrier is positioned like a plug between the carrier and the ampule. ●The LBS020 / LBS060 / LBS180 Rapid Biological Indicator, when used with an automatic reader (incubation temperature 56-60°C), provides a fluorescence test result in a short period (LBS020 – 20 minutes; LBS060 – 60 minutes; LBS180-180 minutes) with an accuracy of ≥97%. Continuing the incubation for 48 hours will yield biological culture results. ●The label on the outside of the plastic pipe contains a chemical indicator strip. |
Population: | ≥1.0×106 per carrier |
Read-Out Time: | No evidence of contamination present in sufficient numbers to adversely affect the finished product. |
Resistance: | D value at 121 ℃ ± 1 ℃ ≥1.5 mins. Z value ≥ 6 ℃. |
Indications for Use: | Used to monitor the effectiveness of Steam Sterilization Processes ranging from 132°C to 135°C. |